Technical Guide Orbitrap Fusion, ETD TECHNICAL INFORMATION

Orbitrap Tribrid Series 3.3 Fix: Instructions are available to fix an issue with the ETD Reagent intensity on Orbitrap Fusion (only) systems.

Orbitrap Tribrid Series 3.3 ETD Fix Instructions

Thermo Fisher Scientific


Orbitrap Fusion (only) ETD systems may show low (~2x) ETD reagent anion signal intensity due to a narrow tuning range of trap lens 1 (TL1), which can lead to longer than necessary reagent injection times and reduced kinetics. The reagent intensity can be recovered with manual tuning, and…

Read more... Download file Email This Item